Automatic tissue processors

AT600  is an automatic tissue processor that perfectly combines approved technology with modern, digital control. Technical and operational characteristics in comparison with earlier released devices are considerably expanded.

The first tissue processor working in the absence of the electric power till 7 o'clock

The  automatic tissue processor are issued in the following modifications: AT600-EKA, AT600-T4EKA, AT1010-EKA, AT2015-EKA

 

Automatic tissue processor AT600-EKA
        Automatic tissue processor AT600T4-EKA           
Automatic tissue processor AT1010-EKA      
Automatic tissue processor AT2015-EKA

                                                                      

TECHNICAL CHARACTERISTICS

 

 

   

DESCRIPTION

AT600-EKA

AT600-T4EKA

 AT1010-EKA AT2015-EKA

 Nominal voltage, V

220±10%

220±10%

 220±10% 220±10% 

 Nominal frequency, Hz

50-60

50-60

 50-60  50-60 

 Rated capacity, W

120

150

 150  150 

 Number of programs kept in memory

6

6

 6  6 

 Quantity of used baskets, pcs.

1

2

 1  1 

 Total of used vessels, pcs.

11

12

 11  11

 Quantity of glass vessels for reagents, pcs.

8

9

 8  8

 Volume of glass vessels, liters

0.6

0.6

 1.0 2.0 

 Capacity tissue basket, cassettes

 15  30  50 100

 Number of thermostats for paraffin

3

4

 3  3

 Temperature in thermostats regulated, °C

30…70

    30…70          

 30…70 30…70 

 Overall dimensions, mm

?560x530-630

?560x530-630   

 ?650x530-630 ?780x530-630 

 Operating time in an autonomous mode, hours, to

6

5

 10 10 

 Mass of the device net, kg.

47

49

 42  45

 

 

    

                         Screen 1. Main screen                                                      Screen 2.  Main menu  
 

       

                                                                  Screen 3.  Setting parameters

On the screen the operator programs for each position:

- processing time from 0 to 999 minutes;

- time of lag of a basket over a vessel, for ensuring running off of a reactant, from 0 to 60 sec.;

- number of shaking a basket during stay in a vessel from 0 to 60 times a minute;

- temperature in thermostats from 30 °C to 70 °C. For each program temperature of thermostats is set separately. 

    

After setting parameters the operator back to screen 2 and go to directive screen 4

    

                                                                                    Screen 4. Directive screen 

On the touch  screen the operator chooses one of the programme from 1 to 6.  The "START" button  carries out start of the chosen program.

 In operating time of the program it is displayed:

- number of a position (vessel) in which there is a processing;

- time before the end of processing in the current vessel;

- the general time before the program termination;

- actual temperature of thermostats (pos. No 9, No. 10, No. 11)

In case temperature thermostats is not correctly, you should adjust temperature on screen 6

                                                             Screen 6. Setting and Adjust parameters

The device has work function  «Timer»  - delayed start . In comparison with similar functions in devices of other manufacturers, time of the delayed start not limited. For setting the timer it is enough to set the date and time switch program.  After installation to press a key - the indicator of red color becomes green and on the touch screen near hours in a periodic mode there will be an inscription the TIME. The programmed  device has a real time clock. Key "setting time" allows the correct time, for example, when the clock is in the winter or summer. 

In case of tension disappearance to alternating current mains, the device instantly passes to power supply from the accumulator.  On the screen in a periodic mode  appears an inscription "NO220V" and the intermittent signal is given. 

 

 

 

 

 

 

 